2. Acupoint Detection
When detecting acupoint, The "timer" should be set to "OFF" and press "detective" at the same time. Insert the penplug into " detective out" socket (A or B), then adjust the "APD" knob until hearing a sound cue. Then, slowly adjust counter-clockwise until a suitable setting is found. Now hold the pen in your hand and place the point on your body until both a signal and light appear which shows the exact location of acupoints.
When detecting acupoint, "APD" is set to any position, there is no indication that shows your skin is too wet. You should dry the skin before detecting. If you regulate the knob to over 7.5, the device will emit a sound cue. At this time, the skin is too dry - wet the skin in order to continue.
During the course of detecting, the correct method is to press the head on the acupoint with slight pressure in a point/touch way. It is also incorrect to touch and press the acupoint for a longer time, for this causes a false reflection point (as cautioned by an unclear sound cue).
Based on the fact that the resistance of certain body points is laser than others, and due to differing skin electrical resistances due to sex, age, health, and season, the electric current used must be adjusted slowly until comfortable.
Caution should be used while detecting. When detecting is complete, "APD" knob should be turned to the "0" position.
3. Galvano-acupuncture
It is considerably efficient for doctors to use this method in for rehabilitative effect. Clip the needle into the patient's acupoint. Select the wave form desired and properly regulate the output strength.
When using this function, the "POWER" is set to "ON" position - set the timer as desired (if more than one hour, timer is set to " ON" position). Press the knob of wave forms selected - a green light will indicate the device is functioning. Then, turn the output switch corresponding to output end
according to need.
4. Hand Massage
Affix the two electric conductible rubber pieces to the place to be massaged (or to the areas of girth). Then, connect the lead wires to the pads, while plugging the other end of the wire into any one of the output sockets in the lower part of the device. Set the "power" is set to "ON" position, the timer as desired (if more than one hour, timer is set to " on " position). Press the knob of wave forms selected - a green light will indicate the device is functioning. Then turn the output switch, corresponding to the output end, gradually from a lower level to a higher level until a tingling sensation is experienced. After a period of treatment, if you feel the current is weak, the output may be increased. If the output is strong yet the feeling remains weak, you may connect two outputs to increase the treatment effect. Several output ends may be used at the same time for enhancing the rehabilitation, such as for sportsmen recovering from fatigue.
5. When finished
the "timer" switch is set to "OFF" position automatically and the "power" switch
should be set to the "OFF" position also. |